Name |
Specification |
Excitation |
Single or two-tone sweep |
Frequency distribution |
Linear or logarithmic |
Frequency resolution |
Better than 1 Hz |
Logarithmic step size |
1/1, 1/3, 1/6, 1/12, 1/24, 1/48, 1/96 |
Number of tones |
> 10000 |
Detector type |
Selective or broadband |
Detector accuracy |
High, medium, low |
Averaging |
Linear, 1 1000 |
Measurement delay |
0-100 s, resolution 0.1 s |
Distortion measurement |
Harmonic, intermodulation, difference frequency |

The purpose of the Amplitude Sweep is to measure the steady-state input/output relationship for an excitation signal having fixed frequency and variable (stepped) amplitude.
Two types of distortion measurement are available in this mode: Harmonic and intermodulation.
Name |
Specification |
Excitation |
Single or two-tone sweep |
Frequency distribution |
Fixed |
Amplitude distribution |
Linear or logarithmic |
Number of amplitude steps |
> 10000 |
Detector type |
Selective or broadband |
Detector accuracy |
High, medium, low |
Averaging |
Linear, 1 1000 |
Measurement delay |
0-100 s, resolution 0.1 s |
Distortion measurement |
Harmonic, intermodulation, difference frequency |
In this mode, the multi-tone excitation signal is used to optimize the execution speed of the steady state frequency response measurement. The multi-tone stimuli can be programmed to comprise a sequence of multi-tone signals, each such multi-tone signal containing a sequence of multi-tones. A given signal of multi-tones may contain an arbitrary number of tones, including the start and stop frequency, as well as the frequency step type (linear or logarithmic). Programming of user defined, arbitrarily distributed frequencies, is supported.
High speed and high signal-to-noise ratio can be obtained simultaneously, by dividing the multi-tone into separate excitation signals. The separate responses are subsequently combined to obtain the final frequency response curve. The phase of each tone in a given signal is randomly selected, thereby lowering the crest-factor of the signal and preventing phase shift in the signal path to cause unexpected overloads.
The multi-tone signal can be highly complex and must be created carefully. The following issues must be taken into account:
Because of the need to measure performance at different frequencies, conventional testing uses a stepped frequency signal to measure frequency response, distortion and noise at many frequency points. However a different class of test signals well suited for this application is the multi-tone signal. Because all data points for all measurements are acquired simultaneously, the test speed is increased dramatically. Other benefits are that the multi-tone signal is a more realistic stimuli than the pure sine wave.
Programming of user defined, arbitrarily distributed frequencies is supported. High speed and high signal-to-noise ratio can be obtained simultaneously, by dividing the multi-tone into separate excitation signals. The separate responses are subsequently combined to obtain the final frequency response curve. The multi-tone stimuli can be programmed to comprise a sequence of multi-tone signals, each such multi-tone signal containing a sequence of multi-tones. A given signal of multi-tones may contain an arbitrary number of tones, including the start and stop frequency, as well as the frequency step type (linear or logarithmic). The phase of each tone in a given signal is randomly selected, thereby lowering the crest-factor of the signal and thus preventing phase shift in the signal path to cause unexpected overloads.
The creation of a multi-tone signal can be rather time consuming and depends on the complexity of the defined signal. To improve the speed the system includes a cash function that stores up to 10 generated signals. A control called index is used for specifying where to store the signal. The function works automatically so before generating the signal it checks if the signal already exist in the cash. Therefore it can also be noticed that the first time the test is performed, it can take several seconds before the actually measurement starts, this is the time used for calculating the signal. The following executions then runs at full speed.
The multi-tone signal can be calibrated to the stored calibration curve. This is selected under the Output group of parameters.
To minimize the effect of intermodulation distortion and other unwanted results, the multi-tone test mode has the feature of automatically dividing the signal into separate signals that later will be combined into a single results. This means that it is hidden for the user that the measurement actually is a result of several separate measurements. This is controlled by the Signals control on the front panel, where it is defined how many groups the signal should be split into. Setting Signals to one is the same as disable this feature. How the signal actually is split is controlled by the Audio Subsystem and optimized with some advanced algorithms.
Name |
Specification |
Excitation |
The multi-tone can be divided into a maximum of 10 signals |
Frequency resolution |
From 1 to 1000 Hz in a 1-2-5 sequence |
Frequency distribution |
Linear, logarithmic or user-defined |
Pre-emphasize |
None, 1. Order lowpass, 1. Order highpass, user-defined |
Amplitude scaling |
Multi-tone RMS, Multi-tone peak, or single tone RMS |
Averaging |
Linear, 1 1000 |
Measurement delay |
0-100 s, resolution 0.1 s |
Distortion measurement |
None |
This mode is used to estimate a systems impulse response by measurement of both the input signal and the output signal. Three excitation signals are available: Impulse, MLS, and Multi-tone.
Name |
Specification |
Excitation |
Pulse, MLS, Multi-tone |
Bandwidth |
30,20,8,4,2 kHz |
Record length |
100 100000 |
Pre-emphasize filter |
None, 1. Order lowpass |
Windows |
Gauss, Exponential, Cos, None |
Averaging |
Linear, 1 100 |
Measurement Delay |
0 100 sec, resolution 0.1 sec |
Stimuli Pause |
0 100 sec, resolution 0.1 sec |
In this mode the instrument can used to measure the internal noise generated in the hearing aid, as recommended in the IEC std. 118-0 (1983). The instrument uses FFT-techniques to synthesize an A-weighted third-octave noise spectrum. The number of octave bands is programmable and only limited by the upper measurement bandwidth. It is possible to bypass the A-weighting of the signal. By using the flexible post-processing facilities the Equivalent Input Noise Spectrum can easily be obtained by subtracting a predetermined pure-tonegain response from the measured third-octave spectrum.
Name |
Specification |
Excitation |
None |
Octave resolution |
1/1, 1/3, 1/12, 1/24, 1/48 |
Weighting |
A-weighted |
Number of octave bands |
Programmable |
Averaging |
Linear, 1 1000 |
Measurement delay |
0-100 s, resolution 0.1 sec |
This mode is designed especially for testing Hearing aids with automatic gain controls circuits according to IEC Standard 118-2. The purpose is to determine the dynamic input/output characteristic. This is carried out in three steps. First, the response to a sine wave modulated by a square pulse is measured. Then the envelope to the response is calculated by applying a RMS detector, which converts each sample points in the response to a RMS value. Finally, attack and release time (or recovery time) is detected by using the following technique: A reference level is calculated as the mean value for the 10 rightmost points in the attack or release RMS response. The attack or release time is then found at the first sample point - searching from right to left, which deviate more than a specified number of dB from the reference level.
Name |
Specification |
Programmable amplitude levels |
2 |
High input duration |
10 ms 2 s |
Low input duration |
100 ms 2 s |
Averaging |
Linear, 1 1000 |
Measurement delay |
0 100 s, resolution 0.1 s |
This mode is used to measure the steady-state behavior of hearing aids with automatic gain controls, as recommended in IEC std. 118-2 (1983). The instrument uses a single tone with programmable frequency and amplitude sweep as excitation. Bandpass filtering and statistical methods are used to ensure high accuracy and steady-state conditions. The measurement result are: Steady-state input/output characteristic (AGC-curve) and AGC-limit (alternatively known as AGC-threshold).
Specification for this mode is the same as for the Amplitude Sweep mode.
In Level mode, the instrument continuously measures the output response for one or more input excitation signals, each with programmable frequency and amplitude. Bandpass filtering of the separate tones is used to ensure high accuracy by excluding the distortion products. This mode can be used to adjust (or test) the gain control position.
Specification for this mode is the same as for the Multi-tone.
This versatile mode is intended to be used as an oscilloscope for monitoring a captured signal in time and/or frequency domain. A single tone generator with programmable amplitude and frequency is available as excitation signal. The delay from start of signal generation to signal capture is also programmable, and so is the sampling frequency and the number of samples to be captured.
The following measurement results are available: Time signal, RMS Spectrum, Nth Octave Spectrum, THD, SNR, Total RMS, Distortion, Noise
Name |
Specification |
Excitation |
Single tone, Multi-tone, Wave file |
Bandwidth |
30,20,8,4,2 kHz |
Record length |
100 100000 |
Averaging |
Linear, 1 100 |
Measurement Delay |
0 100 sec, resolution 0.1 sec |
Stimuli Pause |
0 100 sec, resolution 0.1 sec |

Example on setting up the current measurement.
A wide range of general-purpose post-processing facilities is available in the Audio Analyzer. When a measurement is completed, the result is passed from the Audio Analyzer to the Vector subsystem where it can be stored for later use, or scalar measurement result can be derived from the primary result vector. The Vector subsystem sends measurement data to the Calculate subsystem, which facilitates a smoothing subsystem, an advanced limiter subsystem, and a transformation subsystem (spectrum or histogram).
The system has full support for programming the HI-PRO box.
© microLEX Systems A/S 1998. email microlex@microlex.dk , phone: +45 4576 2100, fax: +45 4576 2200. |